EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read our Privacy Policy

Avante NeoGlider Transport Incubator

Avante

Avante NeoGlider Transport Incubator

Web ID: 12110
Condition: Refurbished
Product Number: 7TI2000RS

Need help choosing?

Speak to one of our expertly trained specialists!

Need help Financing?

Check out financing options from OnePlace Capital.

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

Features

  • Use AC power and DC power alternatively, and DC 12V or DC 24V power in an ambulance
  • Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately
  • Can be equipped with >37 C temperature set function
  • Failure alarm indication
  • Double wall hood with side door for pulling out the bassinet
  • Natural air flow humidity
  • Adjustable height

Configurations available with

  • Main body, including the hood, controller, bassinet, internal batter and observation lamp
  • Oxygen cylinders
  • Oxygen supply system
  • Skin temperature sensor
  • IV pole
  • Mattress
  • Adjustable stand
  • >37 C override temperature mode (optional)

Specifications

General

  • AC power supply: AC110V - 240V/50-60Hz
  • DC power supply: DC12V/10A or DC24V/6A
  • Input power: 400VA
  • Air-temperature control range: 25 C - 37 C
  • Override mode: 37 C - 38 C
  • Baby temperature control range: 34 C - 37 C
  • Override mode: 37 C - 37.5 C
  • Skin temperature sensor accuracy: ≤0.3 C
  • Temperature variability: ≤1.0 C
  • Temperature uniformity: ≤1.5 C (mattress is on the horizontal position)
  • Noise inside hood: ≤55dB(A)
  • Environment noise ≤45dB(A)
  • Failure alarm: Over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm
  • Operating at temperature equilibrium: 90 min (Set temp: 36 C | Ambient temperature: 15 C)

Environmental Temperature

  • Operating range: 10 C ~ 30 C
  • Set temperature should be 3 C higher than the environment temperature
  • Ambient air movement rate: ≤1.0 m/s

Package

  • Gross Weight:
    • Main unit: 80 kg
    • Cabinet: 53 kg

Size

  • Main body: 109 cm x 65.5 cm x 76 cm
  • Cabinet: 138 cm x 61 cm x 40 cm