ECG Acquisition
- R/T (real-time) ECG (12 leads)
- AUTO (12 leads)
- RHYTHM (up to 12 leads)
- DISCLOSE (1 to 3 leads)
Keyboard
- Full alphanumeric capability
Touch Screen Display
- 1024 × 768 pixel resolution
- 30.4 cm × 22.8 cm (15-inch diagonal) color liquid crystal touch screen display with backlight
Patient Module
- Action button allows user to take ECG snapshots from the bedside
Signal Processing/Acquisition Sampling Rate
- 1,000 samples per second per electrode/lead. 12 bit A/D conversion provides 5µV resolution.
Auto Frequency Response
- 0.05-150 Hz, 0.15-150 Hz, 0.5-150 Hz
- 0.05-100 Hz, 0.15-100 Hz, 0.5-100 Hz
- 0.05-40 Hz, 0.15-40, 0.5-40 Hz
Rhythm Frequency Response
- 0.05-150 Hz, 0.15-150 Hz, 0.05-100 Hz
- 0.15-100 Hz, 0.05-40 Hz, 0.15-40 Hz
Filters
- AC noise
- Baseline wander
- Artifact
Printer/Printer Resolution
- High-resolution, digital-array printer using thermal-sensitive paper
- 200 dpi (voltage axis) by 500 dpi (time axis)
Report Formats
- 3x4 (1R, 3R)
- 6x2
- Panoramic 12 (Cabrera)
- 12x1
- Rhythm (up to 12 selected leads)
- Extended Measurements
- One Minute Disclose (1 lead)
- Full Disclosure (five minutes, 1 to 3 selected leads)
Battery Operation Capacity
- Typically 50 ECGs and copies on a single charge or 40 minutes of continuous rhythm recording Recharge
- Seven hours in standby mode to >90% capacity (typical)
Network Connection
- 10/100 Base-T IEEE 802.3 Ethernet via RJ45 connector (standard)
- Optional software required for Proxim Range LAN2 7410 CE PC card wireless LAN connection (Wireless PC LAN card not included)
FAX Capability (optional)
- Group 3, Class 1 or 2 fax Modem (optional for USA & Canada)
- V.90, K56flex, enhanced V.34, V.32bis, V.32, V.22bis and below
Barcode Reader (optional)
- Reads Code 39 (standard and full ASCII)
Barcode Reader (optional)
- Reads cards adhering to ISO 7810, 7811-1, -2, -3, -4, -5, and JIS × 6301 and X6302
ECG Storage
- XML File Format
- 150 ECGs to internal flash memory
- 2 to 3 ECGs typical per 1.4 MB floppy disk
- 150 ECGs per 128 MB PCMCIA card (optional)
ECG File Formats
Power and Environment Line Power
- 100-240 Vac, 50/60 Hz, 150 VA max
Environmental Operating Conditions
- 15° to 35 °C (50° to 104°F)
- 15% to 70% relative humidity (non-condensing)
- Up to 4,550 m (15,000 ft.) altitude
Cardiograph Dimensions
- Approx. 45 × 45.8 × 16 cm (17.7 × 18.0 × 6.34 in.)
Cardiograph Weight
- Approx. 13 kg (28 lb.) including accessories
Patient Interface Module
- Remote, microprocessor-controlled module