EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read our Privacy Policy

Avante Infantia YP2 Infant Incubator

Avante

Avante Infantia YP2 Infant Incubator

Web ID: 14082

Need help choosing?

Speak to one of our expertly trained specialists!

Need help Financing?

Check out financing options from OnePlace Capital.

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

•    Two control modes: Air mode and Baby mode
•    The information can be stored, checked and indicated by the LCD display screen
•    Humidity and Oxygen concentration levels display and servo control system
•    ">37'C" temperature set function
•    Air circulation system with outlets in both the front and back
•     X-ray cassette under the bassinet
•    Embedded integrated sensor module, integrated water reservoir in the drawer
•    Electronic tilting to adjust the incline of the bassinet
•    Adjustable Height
•    Multiple failure alarm indications
•    RS-232 connector for data output
•    Low noise DC motor

Specifications

Power requirement: AC220V-230V/50Hz or AC110-120V/50- 60Hz, 1300VA 
Control Mode: Air mode and Baby mode controlled by micro­computer 
Air temperature control range: 25•c-37°C(override mode 37 C-39'C) Baby temperature control range: 34 ·c-37 C(ovemde mode 37 C-38 C) Accuracy of skin temperature sensor: ± o.2·c 
The temperature fluctuation range between incubator temperature and average incubator temperature(under stable temperature condition): ,s:0.5°C 
Temperature uniformity: ,s:0.8°C(mattress is on the horizontal position) Inclination of the bassinet: ± 12° 
Noise inside hood: ,s,45dB(A) 
Failure alarm: over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure 
alarm, sensor box position alarm, tank position alarm, system alarm, etc
Humidity display range: 0%RH-99%RH 
Humidity control range: 0%RH-90%RH 
Accuracy of humidity display: ± 5%RH(environment temperature is 25'C, environment humidity is 45%RH ± 5%RH) 
Indication range of oxygen concentration: 0%-99% 
Control range of oxygen concentration: 20%-60% 
Accuracy of oxygen display concentration: 
± 2%(oxygen concentration sets ,;;25%) 
± 3%(oxygen concentration sets >25%) 
Accuracy of oxygen control concentration: ± 4% Weighing Range: 100-8000g(weigh1ng system) Resolution for weight indication: 1g(weighing system) Accuracy of weigh: ± 1 %(weighing system) 

0

Catalog

Help & Settings

Equipment

See all

Supplies

See all