• Two control modes: Air mode and Baby mode
• The information can be stored, checked and indicated by the LCD display screen
• Humidity and Oxygen concentration levels display and servo control system
• ">37'C" temperature set function
• Air circulation system with outlets in both the front and back
• X-ray cassette under the bassinet
• Embedded integrated sensor module, integrated water reservoir in the drawer
• Electronic tilting to adjust the incline of the bassinet
• Adjustable Height
• Multiple failure alarm indications
• RS-232 connector for data output
• Low noise DC motor
Power requirement: AC220V-230V/50Hz or AC110-120V/50- 60Hz, 1300VA
Control Mode: Air mode and Baby mode controlled by microcomputer
Air temperature control range: 25•c-37°C(override mode 37 C-39'C) Baby temperature control range: 34 ·c-37 C(ovemde mode 37 C-38 C) Accuracy of skin temperature sensor: ± o.2·c
The temperature fluctuation range between incubator temperature and average incubator temperature(under stable temperature condition): ,s:0.5°C
Temperature uniformity: ,s:0.8°C(mattress is on the horizontal position) Inclination of the bassinet: ± 12°
Noise inside hood: ,s,45dB(A)
Failure alarm: over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure
alarm, sensor box position alarm, tank position alarm, system alarm, etc
Humidity display range: 0%RH-99%RH
Humidity control range: 0%RH-90%RH
Accuracy of humidity display: ± 5%RH(environment temperature is 25'C, environment humidity is 45%RH ± 5%RH)
Indication range of oxygen concentration: 0%-99%
Control range of oxygen concentration: 20%-60%
Accuracy of oxygen display concentration:
± 2%(oxygen concentration sets ,;;25%)
± 3%(oxygen concentration sets >25%)
Accuracy of oxygen control concentration: ± 4% Weighing Range: 100-8000g(weigh1ng system) Resolution for weight indication: 1g(weighing system) Accuracy of weigh: ± 1 %(weighing system)