Features:

  • Optional Dual Access Doors – Allow access from either side, expediting care.
  • Thermal Performance – Patented dual air curtains reduce radiant heat loss from the infant by warming the inner hood surface. A bi-directional airflow feature helps to minimize temperature fluctuations when the access doors are opened.
  • Large Mattress and Dual Probes – The C2000 incubator allows for simultaneous care of twins in the same unit, allowing more incubators for more patients.
  • Internal Noise Level less than or equal to 47dBA – Maintains a supportive environment for both patient and caregiver.
  • Optional In-bed Scale – Weigh the baby in a warm environment with minimal disturbance.
  • 24-hour Data Trends – This unit facilitates record keeping for air and skin temperature, oxygen, humidity, heater power, and weight changes.

Specifications

Air-Shields Incubator with VHA Stand

  • Width: 38.0" (99 cm)
  • Depth: 26.3" (67 cm)
  • Height: 52.5" - 60" (133.4 to 152.4 cm)
  • Weight: 198 lbs (90 kg)

Air-Shields Incubator with Fixed Height Stand

  • Width: 46.8" (119 cm)
  • Depth: 26.3" (67 cm)
  • Height: 56" (142 cm)
  • Weight: 175 lbs (79.5 kg)

Hood Specifications

  • Mattress tray size: 16" x 32" (40.6 81 cm)
  • Mattress to hood: 16.25" (41.2 cm)
  • Access panel opening: 11" (28 cm)
  • Standard hood includes: 4 port doors; 2 iris ports; front access panel; 3 left and 3 right tubing grommets-front
  • Mattress tilt: 12 (1 C)
  • Micro air intake filter: 99.9% efficiency
  • Particle size removal: 0.3 micron
  • Air mode control range: ;20 to 39.0C
  • Air mode override: >37.0C
  • Air flow velocity across mattress: <10 cm/sec
  • Skin mode control range: 34.0 to 38.0 C
  • Skin mode override: >37.0 C
  • Temperature rise time (at 22C ambient): <35 minutes
  • CO2 levels: <0.5% when delivered at 750 ml/min at a point 10 cm above the mattress
  • Operating noise level in hood: 47 dBA (49 dBA with Servo Controlled Oxygen)
  • Casters (4 casters, 1 steering): 5" (12.7 cm)
  • RS-232 output
  • Controller with LCD with brightness control
  • Selectable color combinations: white on blue background (default) or yellow on black background
  • Dual-skin temperature monitoring
  • 6 AC (Mains) outlet
  • Keypad lock
  • 24 Hour trend of: Air temperature, skin temperature (1 and 2), Relative humidity, Oxygen concentration, Heater power

Environmental Requirements

  • Operating temperature range: 20C to 30C ambient
  • Storage temperature range: -25C to 60C ambient
  • Operating humidity range: 5% RH to 99% RH, non condensing
  • Storage humidity range: 0% to 99% RH, non condensing
  • Operating RH sensor temp. range: 20C to 41C
  • Operating O2 sensor temp. range: 20C to 41C
  • O2 sensor calibration pressure: 600 to 900 mm Hg

Alarms

  • Temperature (limiting ≤): 38C for air temp. < 37C; 40.0 C for air temp. > 37 C
  • Baby set temperature: 1 C (or 0.5 C)
  • Air set temperature: +1.5 C/-2.5 C

High/Low Skin Temperature Alarms

  • High/low air temperature
  • High temperature cut-out
  • Air probe failure
  • Skin probe failure/disconnect
  • Low air flow
  • Low humidity (humidity option)
  • Low/high oxygen (3%) (oxygen option)
  • Oxygen calibration required (oxygen option)
  • Oxygen cell difference (oxygen option)
  • Oxygen solenoid failure (oxygen option)
  • Scale disconnect (scale option)
  • Too much weight (scale option)

System Alarms

  • Controller failure
  • Sensor module failure
  • sensor module out of position
  • Motor failure
  • Power failure
  • Heater failure
  • Sensor disconnected
  • Key stuck

Servo Humidity Option

  • Water capacity: 1000ml
  • Refill time: >24 hrs @ 85 RH%
  • Range: 30 to 95%
  • Display accuracy: 6% RH%

Servo Oxygen Option

  • Accuracy: 100% cal 3%
  • Accuracy: 21% cal 5%
  • Range: 21 to 65%
  • Resolution: 1%

Scale Option

  • Weight range: 0 to 15.4 lbs (0 to 7 kg)
  • Weight display resolution: 1 g or 1 oz (OIML = 10 g or 1 oz)
  • Weight accuracy: 2 g 1/2 digit up to 2 kg (OIML = 10 g); 5 g 1/2 digit over 2 kg
  • Weight gain or loss trended over 7 days

Other Options (Attach to Accessory Rail)

  • IV Pole weight limit: 10.4 lb (4.72 kg)
  • Monitor shelf weight limit: 25 lb (11.4 kg)
  • Rear access door: Provided with 2 port doors and 4 additional hood grommets 2 left and 2 right
  • Rear access door: Provided with 2 port doors and 4 additional hood grommets 2 left and 2 right
  • Oxygen tank mount

Drawer Options

  • Large deep swivel type
  • Two shallow swivel out type