Display
- Type: High resolution color TFT display
- Size: 12.1” (diagonally)
- Resolution: 800 x 600 pixels, excellent viewing from any angle
- Parameters: ECG, Respiration, SpO2, Dual NIBP, Temperature & CO2
- Trace Speed:
- 12.5, 25 & 50 mm/sec. for ECG & SpO2
- 6.5, 12 & 25 mm/sec. for Respiration & CO2
Optical Encoder / Keys
- Optical Encoder: Rapid access to all the functions and setting of parameters through single knob
- Hot Keys: Quick action hot keys for alarm acknowledge, NIBP start/stop, recorder start/stop, freeze, stand by, go-to and return to the main screen
ECG Monitoring
- Lead: 3/5 lead with optional ST & Arrhythmia
- Protection: In-built cautery & defibrillator protection
- Pacer Detection: Possible
- HR Range: 20 to 350 bpm (±2 bpm or 2%, whichever is greater)
Respiration (from ECG)
- Principle: Impedance pneumography
- PR Range: 4 to 150 bpm
Pulse Oximetry ( Nellcor Module)
- Tone Variation: Tone variation with change in SpO2 values
- SpO2 Range: 1 - 100%
- Accuracy:
- Large Animal: 100 - 70% (±2 digits)
- Medium Animal: 0 - 69% (unspecified)
- Small Animal: 100 - 70% (±3 digits)
- Perfusion Indicator: Bar graph showing signal strength
- Modes: Normal response & fast response
- Alarms:
- Low: 0 - 95%
- High: 5 - 100%
- PR Range 20 to 230 bpm
Non-Invasive Blood Pressure (Suntech Module)
- Principle: Oscillometic
- Display: Systolic, Diastolic & Mean pressure
- Modes: Manual, STAT (continuous 5 min. operation) and automatic (time interval 2-90 min. selectable)
- Alarm Limits: Selection possible of r Systolic & Diastolic
- Range: 20 - 250 mmHg
- Accuracy: ±3 mmHg
Invasive Blood Pressure
- Measurement Range: -50 ~ 300mmHg
- Channel: 2 channels
- Pressure Transducer: sensitivity, 5μV/V/mmHg
- Impedance Range: 300 ~ 3000Ω
- Transducer Sites: ART, PA,CVP, RAP, LAP, ICP
- Unit: mmHg/kPa selectable
- Resolution: 1mmHg
- Accurancy: ±1mmHg or ±2%, whichever is greater
- Alarm Range: -10 ~ 300mmHg
Temperature
- Channel: Dual channel
- Range: 0ºC to 50ºC (±0.2ºC or -0.4ºF)
- Unit: ºC or ºF
- Probe compatible with YSI 400 series probes (optional)
Capnography (Oridion Module)
Microstream Capnography with display of CO2 waveform and digital values of EtCO2 , FiCO2 & RR. CO2 gets priority for calculating Respiration rate.
- EtCO2 Range: 0 - 99 mmHg
- FiCO2 Range: 0 - 20 mmHg
- Gain: 20, 40, 60, 80 mmHg
- Flow Rate: 50 ml/min. (-7.5 / +15 ml/min.)
- Warm-up time: 40 sec. (typically), Units: mmHg / Kpa / Vol%
- RR Range: 0 - 150 bpm
Alarms
Alarm setting for all parameters.
- Patient Alarm: RED flashing with audio and message indicator
- Equipment Alarm: YELLOW flashing with audio and message indicator
- Alarm Suspend: Continuous YELLOW
- Auto Setting: HR / PR, SpO2, RR, Temperature & CO2
- Manual Setting: HR / PR, SpO2, Systolic, Diastolic, RR, Temperature & CO2
- Re-Alarm Facility: 3 min. for acknowledge alarm
Trends
72 hrs / 100 patients graphical and tabular trend for HR / PR, SpO2, RR, Temperature & CO2 & ST.
- NIBP Trend: Last 240 readings (tabular)
- Alarm Recall: Last 12 critical alarms with date & time
Recorder (optional)
- Type: In-built dual channel thermal array recorder for recording ECG, plethysmograph, Respiration & CO2 waveforms (any two-user selectable) with numerical data, NIBP & Tabular trend recording
- Modes: Direct / delayed / continuous and record on alarm features
Slave Display
- Dimension: Possible with external monitor
- HxWxD: 235 x 260 x 158 mm
- Weight: 4.5 kg. (with in-built battery & thermal array recorder)
Power Supply
- Universal Poser: 95 - 264V AC, 50 160 Hz (±5%)
- Current Rating: 0.7 A (max)
- Battery Type: 14.8V, (4Ah / 8Ah) Lithium ion (in-built)
- Battery Operation: Rechargeable battery for 2 - 4 hrs of continuous operation under full charge condition (typically)
- Charging Time: 8 hrs (typically)