Details

Measures ECG (3/5 lead), Respiration, SpO2 (Nellcor Oximax®), Non-Invasive Blood Pressure (Suntech®), Dual Temperature and Dual Invasive Blood Pressures.

Also available with EtCO2 (Oridion®) and 
Anesthetic Agent monitoring with Automatic Agent ID
 

Features:

  • Simultaneous multi-lead ECG monitoring
  • 72 hour / 100 patients trend
  • Large, medium, and small animal modes
  • Paperless USB data management
  • User selectable screen formats and colors
  • Variance in SpO2 pulse tones
  • Remote control
  • Built-in recorder
  • Li-Ion battery
  • Available with four IPB channels

Wireless Remote:

  • Control the DRE ASM 6000 ProVet from anywhere in the operating room
  • Operates as far as 20 feet from the monitor

Specifications

Display

  • Type: High resolution color TFT display
  • Size: 12.1” (diagonally)
  • Resolution: 800 x 600 pixels, excellent viewing from any angle
  • Parameters: ECG, Respiration, SpO2, Dual NIBP, Temperature & CO2
  • Trace Speed:
    • 12.5, 25 & 50 mm/sec. for ECG & SpO2
    • 6.5, 12 & 25 mm/sec. for Respiration & CO2

Optical Encoder / Keys

  • Optical Encoder: Rapid access to all the functions and setting of parameters through single knob
  • Hot Keys: Quick action hot keys for alarm acknowledge, NIBP start/stop, recorder start/stop, freeze, stand by, go-to and return to the main screen

ECG Monitoring

  • Lead: 3/5 lead with optional ST & Arrhythmia
  • Protection: In-built cautery & defibrillator protection
  • Pacer Detection: Possible
  • HR Range: 20 to 350 bpm (±2 bpm or 2%, whichever is greater)

Respiration (from ECG)

  • Principle: Impedance pneumography
  • PR Range: 4 to 150 bpm

Pulse Oximetry ( Nellcor Module)

  • Tone Variation: Tone variation with change in SpO2 values
  • SpO2 Range: 1 - 100%
  • Accuracy:
    • Large Animal: 100 - 70% (±2 digits)
    • Medium Animal: 0 - 69% (unspecified)
    • Small Animal: 100 - 70% (±3 digits)
    • Perfusion Indicator: Bar graph showing signal strength
    • Modes: Normal response & fast response
    • Alarms:
      • Low: 0 - 95%
      • High: 5 - 100%
    • PR Range 20 to 230 bpm

 
Non-Invasive Blood Pressure (Suntech Module)

  • Principle: Oscillometic
  • Display: Systolic, Diastolic & Mean pressure
  • Modes: Manual, STAT (continuous 5 min. operation) and automatic (time interval 2-90 min. selectable)
  • Alarm Limits: Selection possible of r Systolic & Diastolic
  • Range: 20 - 250 mmHg
  • Accuracy: ±3 mmHg

Invasive Blood Pressure

  • Measurement Range: -50 ~ 300mmHg
  • Channel: 2 channels
  • Pressure Transducer: sensitivity, 5μV/V/mmHg
  • Impedance Range: 300 ~ 3000Ω
  • Transducer Sites: ART, PA,CVP, RAP, LAP, ICP
  • Unit: mmHg/kPa selectable
  • Resolution: 1mmHg
  • Accurancy: ±1mmHg or ±2%, whichever is greater
  • Alarm Range: -10 ~ 300mmHg

Temperature

  • Channel: Dual channel
  • Range: 0ºC to 50ºC (±0.2ºC or -0.4ºF)
  • Unit: ºC or ºF
  • Probe compatible with YSI 400 series probes (optional)

Capnography (Oridion Module) 
Microstream Capnography with display of CO2 waveform and digital values of EtCO2 , FiCO2 & RR. CO2 gets priority for calculating Respiration rate.

  • EtCO2 Range: 0 - 99 mmHg
  • FiCO2 Range: 0 - 20 mmHg
  • Gain: 20, 40, 60, 80 mmHg
  • Flow Rate: 50 ml/min. (-7.5 / +15 ml/min.)
  • Warm-up time: 40 sec. (typically), Units: mmHg / Kpa / Vol%
  • RR Range: 0 - 150 bpm

Alarms 
Alarm setting for all parameters.

  • Patient Alarm: RED flashing with audio and message indicator
  • Equipment Alarm: YELLOW flashing with audio and message indicator
  • Alarm Suspend: Continuous YELLOW
  • Auto Setting: HR / PR, SpO2, RR, Temperature & CO2
  • Manual Setting: HR / PR, SpO2, Systolic, Diastolic, RR, Temperature & CO2
  • Re-Alarm Facility: 3 min. for acknowledge alarm

Trends 
72 hrs / 100 patients graphical and tabular trend for HR / PR, SpO2, RR, Temperature & CO2 & ST.

  • NIBP Trend: Last 240 readings (tabular)
  • Alarm Recall: Last 12 critical alarms with date & time

Recorder (optional)

  • Type: In-built dual channel thermal array recorder for recording ECG, plethysmograph, Respiration & CO2 waveforms (any two-user selectable) with numerical data, NIBP & Tabular trend recording
  • Modes: Direct / delayed / continuous and record on alarm features

Slave Display

  • Dimension: Possible with external monitor
  • HxWxD: 235 x 260 x 158 mm
  • Weight: 4.5 kg. (with in-built battery & thermal array recorder)

Power Supply

  • Universal Poser: 95 - 264V AC, 50 160 Hz (±5%)
  • Current Rating: 0.7 A (max)
  • Battery Type: 14.8V, (4Ah / 8Ah) Lithium ion (in-built)
  • Battery Operation: Rechargeable battery for 2 - 4 hrs of continuous operation under full charge condition (typically)
  • Charging Time: 8 hrs (typically)