Both the Hallowell 2002 and 2000 ventilators are equipped with dual airway pressure alarms that monitor and sound when the peak inspiratory pressure (PIP) of any breath either exceeds the preset maximum working pressure limit (MWPL) or fails to reach 6 cm H2O as in the case of a disconnect.
A Low Supply Gas Alarm, available as an option, would sound when the supply gas pressure falls below 35 psi, 2.4bar. The normal supply gas operating pressure is 50psi or 3.5bar. It is not until 35psi or 2.4bar that the ventilators delivered volume will begin to drop off proportionally with the warining supply pressure.