The AV900 Ventilator is intended to provide continuous mechanical ventilatory support during anaesthesia. The AV900 is a time-cycled, volume/pressure controlled, and pressure limited ventilator for closed circuit ventilation or for use with a Mapleson D circuit. The ventilator is compliance compensated and has a user selectable option of an inspiratory pause fixed at 25% of the inspiratory time.
Options Available for Model
Spirometry
An integral oxygen monitor to measure oxygen concentration in the breathing circuit
Paediatric bellows assembly
Specifications:
Application: Ventilation for use in anaesthesia
Internal Compliance
Adult bellows: 3 ml/cmH2O (nominal)
Paediatric bellows: 2 ml/cmH2O (nominal)
Minute Volume Range: N/A - this parameter cannot be set.
Tidal Volume Range
Adult bellows: 20 to 1600 ml (±10%)
Paediatric bellows: 20 to 350 ml (±10%)
At ambient temperature of 20oC (±10%) and ambient atmosphere of 101.3 kPa (±10%).
Frequency (Rate) Range: 4 to 100 bpm, limited by tidal volume setting
Inspiratory Phase Time Range: 0.07 to 11.54 seconds (100 bpm, I:E 1:8)
Expiratory Phase Time Range: 0.14 to 13.33 seconds (100 bpm, I:E 1:0.3 and 4 bpm, I:E 1:8)
Inspiratory/Expiratory Phase Time Ratio Range: (I:E ratio) : 1: 0.3 to 1: 8.0- limited by tidal volume and rating.
Pressure Control Range (pressure ventilation mode): 10 to 70 cmH2O (±10%)
Inspiratory Flow Range: 2 to 75 L/ min
Airway Pressure Limit (volume ventilation mode): 10 to 80 cmH2O (±10%)
Inspiratory Triggering: N/A
Inspiratory Triggering- Response Time : N/A
Maximum Safety Pressure: 80 cmH2O
Maximum Working Pressure: 80 cm H2O
Minimum Safety Pressure: -10 cm H2O
Minimum Working Pressure: Atmospheric
Sub-atmospheric Pressure Range: None
Expiratory Resistance: 3 cmH2O (30 L/min)
Sigh Characteristics: None
Inspiratory Mixture No mixture controls
Flowmeters: None
Manual Changeover: None
Inspiratory Relief Valve: Fixed, 80 cmH2O
Humidifiers and Monitors: No humidifier. Volume monitor optional (spirometer). Pressure monitor built in
System Pressure Gauge: None
Airway Pressure Gauge: Airway pressure value is displayed on screen
Spirometer: Optional
Accuracy: ±10%
Minimum detectable: 300 ml Vt minimum in Spontaneous mode.
Range: 20 to 1600 ml in all other operating modes.
Resolution: 1 ml
Maximum Inspiratory Pressure: 0 to 100 cmH2O
Alarms (in priority order)
Vent. inop. (ventilator inoperative)
High airway pressure
High cont. (continuing) airway pressure (PEEP)
Low supply pressure
Low airway pressure
Negative airway pressure
Low tidal volume
High tidal volume
Incorrect rate or ratio
Low battery
Mains failure
Power Source:
Electrical: 100 to 240 VAC ; 50/60 Hz, universal input ; 0.30 A - 0.18 A
Input Drive Gas: Oxygen or air (dry, and oil free) at 38 to 100 psig (262 to 689 kPa).
Power Consumption:
Electrical: 37 watts maximum
Gas: 75 L/min intermittent maximum inspiratory flow.
Dimensions:
Height: 385 mm
Height of control unit only: 150 mm
Width: 240 mm
Depth: 300 mm
Weight:
Weight with adult bellows: 9.0 kg
Weight with paediatric bellows: 8.7 kg
Weight of control unit only: 7.6 kg
Method of Disinfectionor Sterilisation: Bellows base assembly and inside of bellows require sterilisation
Bacterial Filter: None
Fail Safe Mechanism: Battery back-up in case of mains electricity failure
Reliability MTBF: 5 x 106 to 50 x 106 cycles
Waveform Tests: Not applicable
Volume Tests: Not applicable
(A) Mobility: Secure mounting on anaesthesia machine required.
(B) Mounting: Mounting bracket available as optional extra. The bellows assembly can be separated from the control unit for remote mounting.
Fuse (mains supply): Two fuses, 2 A, 240 V rating, 20 mm, antisurge.
Environmental:
Ambient Temperature
Storage: -5 to 50ºC (23 to 122ºF).
Operating: 10 to 38ºC (50 to 100ºF)
Humidity: 10-95% RH (relative humidity), noncondensing
Altitude: Up to 2775 m (9000 feet)
Ingress protection: Conforms to EN 794-1 Clause 44 (spillage test)