Display
Display size and type - 9 in monochrome EL display, Display resolution 640 x 400
Dimensions and weight:
- Monitor (WxDxH) 325 x 160 x 210 mm/12.8 x 6.3 x 8.3 in
- Battery module (WxDxH) 290 x 150 x 40 mm/11.4 x 5.9 x 1.6 in
- Weight w/o battery 4.2 kg/9.3 lb.
- Weight w/ battery 6.1 kg/13.5 lb
General
- Input voltage - 12.5 V DC
- Graphical trends - 2, 4, 12 and 24 h after NIBP measurement
- Alarms - Adjustable high and low alarms for HR,?Resp, NIBP, SpO2, Temp, Invasive Pressures and CO2
- Safety standards -?Designed to meet IEC 60601-1, CAN/ CSA C22.2 No. 601.1-M90 and UL 2601-1, CE- marking according to Directive 93/42/EEC
- I/O connections - RS-232 serial output, Defib. sync. signal, Nurse Call and Local printer
ECG
- Measured leads - I, II or III
- Frequency response - 0.5 to 30 Hz, (-3 dB, with 50 Hz reject filter), 0.5 to 40 Hz (-3 dB, with 60 Hz reject filter)
Pacemaker Pulse Detection
- Detection level - 5 to 500 mV
- Pulse duration - 0.5 to 2 ms Sweep speed (trace) 12.5, 25 or 50 mm/s
- Heart rate/Pulse rate
- Measurement range - 30 to 250 bpm
- Measurement accuracy - ? 5 % or ? 5 bpm whichever is greater
Respiration
- Measuring range - Impedance, 0.2 to 6 ohm
- Max. base impedance - 5 kohm
- Respiration range - 4 to 120 resp/min
- Measurement accuracy - ?5% or ? 5 resp/min whichever is greater
- Sweep speed (trace) - Slow 0.62 mm/s, fast 6.25 mm/s
Pulse Oximetry (SpO2/Pleth)
- Measurement range - 40 to 100%
- Measurement accuracy* - 100 to 80%, ?2 digits (? 1SD), 80 to 50%, ?3 digits (? 1SD), 50 to 40%, unspecified
Non-Invasive Blood Pressure
- Measurement range - adult 25 to 260 mmHg, child 25 to 190 mmHg, infant 15 to 140 mmHg
- Typical measurement time - adult 23 s, infant 20 s
Temperature
- Measurement range - 10 to 45?C (50 to 113?F)
- Measurement accuracy - ?0.1?C (25 to 45.0?C), ?0.2?C (10 to 24.9?C)
- Probe type - Compatible with YSI 400 series
Invasive Blood Pressure (F-LMP1 models only)
- Measurement range - -40 to 320 mmHg
- Measurement accuracy - ?5 % or ? 2 mmHg
- Transducer sensitivity - 5 μV/V/mmHg
- CO2 Mainstream (Optional)
- Measurement range - 0-99 mmHg / 0-13 vol % / 0-13 kPa
- Accuracy** - 0-40 mmHg /0-5.3 vol %: ?2 mmHg /?0.3 vol %
41-76 mmHg /5.3-10 vol %: ?5% of reading
77-99 mmHg /10-13 vol %: ?10% of reading - Respiration range - 4-150 breaths/minute
- Calibration - No routine calibration required
- Mainstream CO2 sensor weight - < 18,5 g (excluding cable)
- Cable length - 3 m
- Warm-up time - < 80 s (from 25?C ambient)
- Response time - 100 ms
Pulse Oximetry, Nellcor SPO2/Pleth (Optional)
- Calibration range - 100 to 70%
- Measurement accuracy*** - ? 3 digits (? 1SD) for sensor types D-YS, DS-100A, OXI-A, OXI-P/I
- Plethysmographic waveform
Power Adapter
- AC voltage range - 220-240 V /100-120 V ~ 50/60 Hz?
- Allowed AC voltage fluctuations - 100 V -10% to 120 V +10%, 220 V -10% to 240 V +10%
- DC output voltage - 12.5 V and 18.5 V