  • The Dinamap Portable Vital Signs Monitor is designed to meet the specific needs of monitoring vital signs in the Medical/Surgical units.
  • Designed to noninvasively and automatically measure systolic and diastolic pressure, and pulse rate for neonatal or adult/pediatric patients.
  • Results are displayed on large, easy-to-read digital displays.
  • Alarm and visual signals indicate when pressure goes outside preset limits.
  • Various cuff sizes are available.
  • DRE offers the Dinamap 8100 / 8100T NIBP monitor refurbished.

Specifications

Specifications:

  • Designed & Manufactured To Meet The Requirements of The Following Specifications: Electromedical Equipment (Domestic Units), Medical and Dental Equipment, Regulations for Sale and Testing of Electrical Equipment, Safety of Medical Electrical Equipment (International Units), British Standards Institution (British Units).

Mechanical Specifications:

  • Size: 9.1"H x 7.9"W (excluding knob) x 6.9"D
  • Weight: 9.6 lbs
  • Mountings:
    • Horizontal Surface: 4 Rubber Feet.
    • Pole Mount: Integral pole mount and clamp accepts 3/8 inch to 1-1/4 inch diameter pole.
  • Power Cable:
    • Domestic: 10 foot, detachable, blue jacketed 16 gauge terminated with a three-prong hospital grade plug.
    • International: 10 foot, detachable, blue jacketed, 16 gauge unterminated.

Electrical Specifications:

  • Input Voltage:
  • 100 Vac/50 Hz - 2 each, 0.5 Amp, 3AG, 250 V.
  • 120 Vac/60 Hz - 2 each, 0.5 Amp, 3AG, 250 V.
  • 220 Vac/50 Hz - 2 each, 0.25 Amp, FST, 250 V.
  • 240 Vac/50 Hz - 2 each, 0.25 Amp, FSt, 250 V.?
  • DC Fuse Requirements: Battery Fuse - 1 each, 2.0 Amp, FST, 250 V.
  • Battery: 12 volt, 2.4 Amp-hour, sealed lead acid

Performance Specifications:

  • Cuff:
    • Adult/Pediatric: 0 mmHg to 250 mmHg
    • Neonatal: 0 mmHg to 235 mmHg
  • Initial Cuff:
    • Adult/Pediatric: 178 15 mmHg
    • Neonatal: 125 15 mmHg
  • Blood Pressure Determination Ranges:
    • Systolic:
      • Adult/Pediatric: 30 to 245 mmHg
      • Neonates: 30 to 190 mmHg
    • Diastolic:
      • Adult/Pediatric: 10 to 210 mmHg
      • Neonates: 10 to 160 mmHg
    • Pulse:
      • Adult/Pediatric: 40 to 200 bpm
      • Neonates: 40 to 220 bpm
  • Blood Pressure Determination Time: 20 - 45 sec. 120 sec. max.
  • Overpressure Cutoff:
    • With Pneumatics PWA 315-292 or -296
      • Adult/Pediatric: 310 mmHg, 45 mmHg
      • Neonates: 235 mmHg, 10 mmHg
    • With Pneumatics PWA 315-317
      • Adult/Pediatric: 300 mmHg, 30 mmHg
      • Neonates: 235 mmHg, 10 mmHg
  • Pulse Rate Accuracy: 3.5 percent
  • Battery Charging: To atleast 90% capacity in 16 hrs.
  • Battery Operating Time: 6 hrs minimum operation