Features:

  • Fluid delivery applications include parenteral fluids, drugs, electrolytes, total parenteral nutrition (TPN) solutions, whole blood, and blood products
  • Automatically loads administration set tubing
  • Can be used with a variety of source containers
  • Micro and Macro rate range
  • Piggybacking Secondary Medications with Callback Option
  • Does programming modes
  • Configurable air sensor settings
  • user selectable downstream occlusion pressure settings (psig and mmHg)
  • Fluid label library
  • Flow check display indicating downstream in-line resistance
  • Panel lockout with optional Auto Lock function minimizes potential for tampering and IV set removal
  • Configurable Pump Personality feature sets
  • Battery charge level indicator
  • Nurse call
  • RS232 communications port
  • Built-in diagnositcs

Configurable Models:

  • Single Channel
  • Triple Channel

Specifications

  • Dimensions: 10.0" H x 7.75" W x 7.75" D | 254 x 197 x 187 mm
  • Weight: 10.4 lb | 4.7 kg
  • Keep Vein Open (KVO) rate: 0.1-5 ml/hr in 0.1 increments or programmed rate, whichever is less
  • Battery: 12 volt, 2.0 Ah sealed lead acid
  • Battery Life: 5 hrs operating time at 100 mL/hr (with fully charged new battery). 2.5 hrs operating time at 100 mL/hr for the warranted life of the battery
  • Battery Recharge: 8 hrs to recharge to 80% capacity
  • AC Power Requirements: 100-120 VAC, 60 Hz | 220-240 VAC, 50/60 Hz, automatically sensed
  • Power Cord: Approx. 2.9m (9ft)
  • Fuses: 1.6 A Type T (Time Delay), 250 V, two provided
  • Range of programmable:
    • Primary Infusion: 0.1 - 1200 mL
      • 0.1 - 99.9 mL in 0.1 mL/hr increments
      • 1-1200 mL in 1 mL/hr increments
    • Piggyback Infusion: 0.1 - 500 mL/hr
      • 0.1 - 99.9 mL in 0.1 mL/hr increments
      • 1 - 500 mL in 1 mL/hr increments
  • Volume to be Infused: 0.1 - 9999 mL
    • 0.1 - 99.9 mL in 0.1 mL increments
    • 1 - 9999 mL in 1 mL increments
  • Priming Rate: 500 mL/hr
  • Power Consumption: 40 watts maximum
  • Accuracy:
    • 5% or better for any one hr period over the course of 72 hrs at a programmed rate of 100 mL/hr
    • For rates below 1 mL/hr, the accuracy can be maintained at 10% or better for any one hr period of 0.5 mL of delivery, whichever is greater
  • Warranty: 6 months