The sidestream module is a great solution for monitoring anesthesia in procedures that involve infants and adults. The plug-in Anesthesia Gas Module expands measurement capabilities for Avante Waveline monitors and upgrades the Waveline Touch into a fully-featured O.R. Monitor. It also adds Automatic Agent ID to the Waveline Pro. For the accurate and continuous measurement of Automatic Agent ID, Anesthetic Agent Concentrations (HAL, ENF, ISO, SEV, DES), CO2, O2, N2O, MAC and Respiratory Rate.
Features:
Weight: 400 g (cable excluded)
Size: 49 x 90 x 100 mm (3.3 x 5.6 x 1.6 inches)
Operating temperature: 5 to 50 C (41 to 122 F)
Storage: -40 to 70 C (-40 to 158 F)
Humidity: 10 - 95 %, non-condensing
Atm. pres.: 525 - 1200 hPa (4,572 m)
Gas Analyzer
Calibration: No routine calibration is required. An automatic zero reference calibration is performed once every startup and once every 8 hours.
Compensation: Pressure, temperature and broadening effects on CO2.
Warm-up time: 20 seconds.
Patient connections
IRMA Airway adapter: Adult/Pediatric: 6 mL dead space
IRMA Airway adapter Infant: 1 mL dead space
Nomoline: Sampling line with luer lock co
During standard conditions:
CO2: Range0-15 vol%; Accuracy: (0.2 vol% + 2 % of reading)
N2O: Range 0-100 vol%; Accuracy: (2 vol% + 2 % of reading)
HAL, ISO, ENF: Range 0-8 vol%; Accuracy: (0.15 vol% + 5 % of reading)
SEV: Range 0-10 vol%; Accuracy: (0.15 vol% + 5 % of reading)
DES: Range 0-22 vol%; Accuracy: (0.15 vol% + 5 % of reading)
O2 Range0-100 vol%; Accuracy: (1 vol% + 2 % of reading)
Rise time: ≤ 250 ms, N2O, Agents ≤ 350 ms, O2 ≤ 450 ms
Total system response time: < 3 seconds
Breath detect: Adaptive threshold, minimum 1 vol% CO2 change
Respiratory rate: 0 - 150 bpm 1 bpm
Certifications: CE marked according to the 93/42/EEC Medical Device Directive