Non-Invasive Blood Pressure
Oscillometric Principle, Display of Systolic, Diastolic and Mean pressure
- Modes: Manual, STAT (5min continuous), Automatic (interval 2 min. to 90 min.)
- Range: 20 to 255mmHg
- Auto Zero: Zero Pressure Reference is automatically established after every reading
- Cuffs: Quick Connect/Disconnect
- Safety Features: Automatic deflation if - pressure exceeds 300 mmHg for adult and 150 mmHg for neonate – Measurement time exceeds 180 seconds (adult) 90 seconds (neonate) – microprocessor failure.
- Alarm limits: Selectable for Systolic and Diastolic
- Range: 20-250+mmHg
- Accuracy: +/-3 mmHg
Pulse Oximetry
Pitch variation with change in oxygen saturation.
- SpO2 range: 0 - 100%
Display of pulse strength, plethysmograph, PR and SpO2 values. Display of perfusion on bar graph - Accuracy: 100 to 70 % (+/- 2 digits ), 69 to 50 % (+/- 3
digits ), 49 to 0 % (unspecified) Accuracy in
neonatal monitoring 100 to 70% (+/- 3 digits) - Modes: Normal and Fast Response
- Alarms: Low 50-95% - High 55-100%
Alarms
Alarm setting for all parameters.
- Patient alarm: RED flashing with audio and message indicator
- Equipment alarm: YELLOW flashing with audio and message
indicator. Alarm auto-set facility provided Auto setting of alarm for PR and SpO2. - Manual settings: Systolic and Diastolic
- Alarm Recall: Last 16 critical alarms with time and date Volume control: 4 levels of volume Trends 24 hr. of Graphical and Tabular trend for HR/PR, SpO2
- NIBP trend: Last 100 readings (tabular)
- Alarm Recall: Last 16 alarm events with date, time and message. 2hr. zoom for graphical trends with mark, record and print facility.
- Time Scale: 12 Hr. (resolution of 3 min) and 2 hr (resolution 30 Sec) for graphical trend Recording Optional in-built dual channel thermal recorder for recording two waveforms (selectable) with numerical data and alarm
status. NIBP and tabular trend recording with patient information. Direct / Delayed / Continuous modes of recording. Record On Alarm facility (selectable) - Speed: 25 mm/sec or 50 mm/sec
Dimensions
- General Dimension: 145 X 220 X 185 mm (H X W X D)
- Weight: 3.5 kg. with in-built battery and recorder.
Power Supply
- Universal power supply: 95-265V AC, 50 / 60 Hz
- Power ratings: 40 Watt
- Battery type: 14.8V, 14.4Ah (Li-ION)
- Operating time: 4 hours on fully charged battery (with full load & NIBP auto mode 2 mins. interval & recorder on for 2 mins.)
- Charging time: 100% charge in 6 hr. for fully discharged battery
- External Battery: 12V DC input
Display
- Resolution: 640 x 480 pixels
- Type: Color
- Size: 6.4”
Back Panel
- VGA output
- Mains power on/Off
- RS-232 Output
- Ground Connection
- External Printer Port
Trace Speed
- 12.5, 25 or 50mm/sec for ECG and Spo2